Ika pipette is not allowed - ika pipette common problem solution

In the process of using the ika pipette, the most common problem encountered by the user is that the pipetting is not accurate. Therefore, Lebe has sorted out this and hopes to help everyone use the pipette.

When we are inaccurate in pipetting, we must first judge ika What is the pipette in the end is not accurate: is it too big or too small? There are two ways to do this. The first is to weigh the pure water with a balance (the specific method refers to the previous article by Lebe), and the second is that the experienced user can judge by observing the position of the liquid surface in the tip (for The same liquid and the same tip are compared with the previous case). When it is judged that the pipetting amount of the pipette is too large or too small, we can analyze the specific problem. Now Lebe is coming to analyze one by one.

one. Large amount of pipetting

In general, there are not many cases where the amount of pipetting is too large, but if it is encountered, it is generally caused by the following two reasons: First, the operation is wrong. That is, the suction/discharge button is pressed to the end when sucking, but the actual only needs to press the first gear (the first stopping point); the second is that the ika pipette is not allowed, and the special screw is needed to adjust the quantitative screw ( This situation is best left to the supplier to resolve).

two. Small amount of pipetting

The small amount of pipetting is the most common problem that is easily encountered during the use of ika pipettes. There are many reasons for this. The following Laibei will give a judgment order, hoping to help users solve ika pipetting. Problems with the use of the device.

The first step is to check the tips. First check if the tip is installed and pay attention to whether the tip is deformed. You can also reinstall a new tip with no abnormal appearance to confirm whether the pipetting is still small. If the condition of the small amount of pipetting still does not change, then the reason for eliminating the tip enters the second step;

In the second step, check the handle. Check whether the joint between the sleeve and the tip is smooth. If there is a slight scratch or rough feeling caused by friction with the tip for a long time, it is recommended to replace the sleeve (if you need to replace the sleeve, please let the supplier handle it, let The supplier also checks the piston system of the ika pipette while changing the sleeve handle; check if the sleeve is blocked (when the amount of liquid suction is very small or even can not absorb liquid at all, it is likely to be blocked), the specific operation: Lower sleeve (must be careful when removing the handle, avoid missing the small parts on the piston system), the larger diameter end of the sleeve is facing your own eyes, and the smaller diameter is facing the direction of light, if not in the handle If you see a bright spot inside, the handle is blocked. We can use a slender wire to clear the handle. After confirming that there is no abnormality in the sleeve, confirm whether the condition of the small amount of liquid migration still exists. If there is still no improvement, proceed to the third step;

In the third step, check the airtightness of the ika pipette. Set the range of the ika pipette to the maximum, and then assemble the tip: (1) Pipette with a maximum range of 200 ul and above. Draw pure water and hang it vertically for more than 15 seconds. See if the tip of the tip has Water droplets form and even drip, if there is no change, it means good air tightness; (2) Pipette with a maximum range of 200ul or less, after sucking pure water, keep the tip of the tip under the water surface, stay for more than 15 seconds, see the inside of the tip Whether the liquid level has dropped, if there is no drop, it indicates that the air tightness is good. If the airtightness of the pipette is found to be good after the inspection, and the pipetting amount is still small, it is recommended to adjust the dosing screw to the supplier; if the airtightness is not good, proceed to the fourth step;

The fourth step is to check the piston system. After removing the piston, check whether the piston is contaminated. (See the previous article by Lebee, “How to remove the pipette”.) Clean the piston and apply the oil and then install the piston to check if the pipetting amount is too small. If you still can't solve the problem, please hand it over to the supplier; if you find that the seal is aging or the piston is rusting and other serious problems, please also hand it to the supplier for disposal. It is generally not recommended that users purchase the parts themselves and then replace them themselves. Lebe recommends that they be handled by a professional engineer.