Torque wrench tester operation method, many users use the torque wrench tester often do not like to read the detailed instructions, are bothersome, but if the torque wrench tester is not used properly, it will greatly affect your work efficiency, It also shortens the service life of the torque wrench calibrator. If you want to improve your work efficiency and lengthen the use time of the torque wrench tester, then let's learn how to operate the torque wrench tester. Torque Wrench Tester: Torque wrench tester picture Torque wrench tester operation method: Portable Torque Wrench Tester Picture The above is the operation procedure of the torque wrench tester brought by the small dry torque wrench tester. It is hoped that the user of the torque wrench tester will practice after seeing it, so as to improve the work efficiency and extend the life of the torque wrench tester. . Read this article also read: Torque tester industry needs stable development Digital torque tester repair method Hybrid Stepper Motors,Brushless DC Motors,Linear Motor,Gear Motor,Step-Servo Motors,Brushless Motors Changzhou Satidi Import and Export Co., Ltd. , https://www.guanjiejt.com
1. Before using the torque wrench tester, check if the power of the instrument is sufficient. If the battery is insufficient, please charge it first (it can also be used when charging).
2. Place the square probe of the test object into the bearing square hole of the tester, and use the handle at the other end of the test object above the height adjustment block (if the length is not enough, the whole load device can be adjusted freely) .
3. Turn on the power switch and the displayed torque value is zero. If it is not zero, press the clear button to clear the torque value.
4, before the test, * need to set the upper and lower limits, zui small storage value, zui small peak hold value, automatic peak time, automatic shutdown time, gravity acceleration and so on.
1. Upper limit setting: The user sets the upper limit value and can freely set according to the need. The upper limit value is the automatic sound and light alarm, and the upper limit is not higher than the full scale.
2. Lower limit setting: The user sets the lower limit value and can be freely set according to the need. The lower limit value is the automatic light alarm, and the lower limit value must not be higher than the set upper limit value.
3, zui small storage value setting: the user sets the small storage value of zui according to the storage needs, the data smaller than the value will not be stored.
4, Zui small peak hold value setting: the user according to the peak, automatic peak measurement needs to be freely set, data less than this value is not saved by the peak.
5. Automatic peak time setting: The user can freely set the time required to maintain the peak value according to the automatic peak measurement state from 1 second to 99 seconds.
6. Automatic shutdown time setting: In the no-operation state, the automatic shutdown time can be set freely from 0 minutes to 9 minutes (when set to 0 minutes, it means to cancel the automatic shutdown).
Torque wrench tester operation method