How to use and maintain the adjustable high speed homogenizer

How to use the adjustable high-speed homogenizer and how to use it for maintenance, the following will be explained in detail by the Shanghai Bashu technicians, I hope to help you.

An adjustable high-speed homogenizer to use

1, the adjustable high-speed homogenizer is first placed on a level work bench, (with a hex key) required for the installation tool must be homogenized in place first with a cup filled with water or vessel when power test machine , so that the plum blossom head is submerged in the water, turn on the power switch, adjust the speed knob slowly and quickly, no obvious metal friction sound ( if there is metal friction sound, turn off the switch, loose expenditure tool 6m / m support nail will be outside The fixed sleeve is randomly turned to the upper position and then tightened ) . Can work after finishing.

2. Cut the plants or organisms that need to be tested into small pieces or small pieces, place them in a glassware that is randomly equipped, and sandwich them in the middle of the magnet holder ( seven-shaped shape ) . Loosen the fixed motor base 8m/m and tighten the nails. Align the mouth of the vessel downwards so that the homogenization head is 2-3m/m from the bottom of the vessel , and then tighten the 8m/m nail. The working time should not exceed 5 minutes.

High speed homogenizer

Second, the use of homogenizer precautions

1, the arbor and the first shaft with good spring characteristics of the element will move downwardly to the foot connector groove (stepped foot to touch down firmly) machine to do the operation.

2, before the start, pay attention to whether the motor supply voltage on the match.

3 , carbon brush often check, if it is too short to replace the new carbon brush, if you find that the carbon brush spark is not normal, you should stop using it immediately, check the cause of the obstacle, repair it before use, in order to avoid the rectifier and coil burned out.

4 , due to the high motor speed, if the continuous use will burn the motor, so the use time is limited to three minutes, stop for 5 minutes before use.

5 , the glass device should be aligned with the center axis of the machine, centered, no shaking on all sides, before starting.

6 , the machine should not be idling, must use a small amount of liquid or grease when using.

7, when put into the material to be added slowly, slowly open the first file, the file open fast, but slow speed only for starting role, should not be used.

Third, adjustable high-speed homogenizer maintenance

1. After long-term use, the polyvinyl fluoride in the high-speed homogenizer tool is a worn part for users to replace.

2. After each work, remove the tool and rinse it with clean water.

3. Cleaning method: first insert the dispersing head into the polyvinyl fluoride. At this time, the dispersing head step is 0.2 mm longer than the sleeve length . If the sleeve is too long, the end of the small head can be used to reduce the size, and then the polyvinyl fluoride is used. The sleeve is pressed into the fixed head by hand, and then the dispersing head is inserted into the sleeve. At this time, the dispersing head must be freely rotatable in the sleeve. If the sleeve is tight, the φ 5 drill bit can be inserted into the sleeve and pulled. The dispersing head is inserted into the sleeve and fastened to the small shaft, and finally the fixing sleeve and the fixing head are fastened.